ATI Orion Model 130 Conductivity Meter

Description:

Location: E138

Please refer to the owner's manual for further information.

Specifications / Capabilities:

Measurement Modes
Conductivity Range Resolution
0.0 to 199.9 µS cm-1
0 to 1999 µS cm-1
0.00 to 19.99 µS cm-1
0.0 to 199.9 mS cm-1
0 to 500 mS cm-1
0.1 µS cm-1
1 µS cm-1
0.01 µS cm-1
0.1 mS cm-1
1 mS cm-1
Accuracy ±0.5% of measurement value ± 1 digit at operating temperature -10 °C to +55 °C
Salinity
Range 0.0 to 70.0
Resolution 0.1
Accuracy ± 0.1 in the range 0.0 to 42.0 test sample temperature 5 °C to 25 °C
± 0.2 in the range 0.0 to 42.02 at test sample temperature 0 °C to 5 °C
better than 3% of measurement value in the range 42.0 to 70.0
TDS
Range 0 to 1999 mg-l-1
Resolution 1 mg-l-1
Accuracy ± 2 % ± 1 digit at test sample temperature 15 °C to 35 °C
Temperature
Range -5 to 99.9 °C
Resolution 0.1 °C
Accuracy 0.1 °C ± 1 digit at operating temperature -10 °C to +55 °C
Cell constant Setting ranges: 0.1 cm-1 and 0.475 cm-1 fixed
0.45 cm-1 to 1.30 cm-1 adjustable
Reference Temperature 20 °C or 25 °C selectable


ATI Orion 130 Conductivity Meter