Controller Operation (Eurotherm 818)
- The NIST CNR instrument control program is set-up for remote setpoint changes with proper ramp rates automatically invoked. Some important front-panel operations are outlined below
- "SHORT SCROLL" refers to quick presses of the scroll button which toggle the lower display between SP and OP (set point and output) status.
- "LONG SCROLL" refers to pressing and holding until the full parameter set {Pr, SP1, SP2, ... , Sbr } begins to cycle. Note that a flashing dot indicates that a parameter may be changed using the up/down arrow keys.
|
- Set Point
- SHORT SCROLL (if necessary) to SP mode , press arrow keys to change value (hold arrow key to accelerate change)
- Output Limit
- LONG SCROLL to the "Hi" parameter, then immediately press the up or down arrow once to display present setting [% full scale]
- Press up/down arrows as needed to adjust setting
- Press scroll to accept value (press repeatedly to scroll back to the default display, or wait ~60 seconds)
Refer to the table at the right for proper settings.  When heating from room temperature, increase the hi setting gradually (i.e. DO NOT SUDDENLY DUMP MAXIMUM CURRENT INTO THE FURNACE)
|
|
Tmax [C] |
|
"hi" setting |
|
500 |
|
15 |
|
700 |
|
18 |
|
900 |
|
23 |
|
1100 |
|
29 |
|
1300 |
|
38 |
|
1500 |
|
50 |
|
- Displaying & Changing the Ramp Rate
- LONG SCROLL to "Pr", press arrow to display present value (rate in degrees/hour), press arrow again to adjust, press scroll to accept, continue scrolling until you return to the default display
- Note: The short scroll parameter "SP" is equivalent to the long scroll parameter "SP1". When ramp is active, SP/SP1 is the ramp target or endpoint. Otherwise, SP/SP1 is the static setpoint.
- Activating a Ramp
- Press RUN/HOLD key until the display indicates RAMP (this changes to "E" when ramp is complete). The setpoint will jump to the present temperature and immediately begin ramping to the target (SP/SP1).
|