| Guide | NG5, 50 mm (width) × 120 mm (height) |
| λ | from 5 Å to 12 Å, Δλ/λ 20 % |
| Flux at Sample Position |
2.0 × 106 neutrons/cm2/s at 8 Å |
| Typical sample size |
3 cm × 3 cm |
| Accessible scattering angle range |
from 2° to 105° |
| Accessible Q range |
from 0.050 Å-1 to ≈1.65 Å-1 at λ = 6 Å |
| from 0.035 Å-1 to ≈1.20 Å-1 at λ = 8 Å |
| from 0.025 Å-1 to ≈0.80 Å-1 at λ = 12 Å |
| Maximum field integral | 0.44 T·m |
| Dynamic range |
from 5 ps to 15 ns at λ = 6 Å |
| from 10 ps to 40 ns at λ = 8 Å |
| from 35 ps to 100 ns at λ = 12 Å |
| Detector | 32 cm × 32 cm PSD |