Guide | NG-A, 50 mm (width) 70 mm (height) |
λ | from 4.5 Å to 12 Å, Δλ/λ 20 % |
Flux at Sample Position |
3.1 × 107 neutrons/cm2/s at 5 Å |
| 2.7 × 107 neutrons/cm2/s at 6 Å |
| 1.3 × 107 neutrons/cm2/s at 8 Å |
| 5.0 × 106 neutrons/cm2/s at 11 Å |
| 1.4 × 106 neutrons/cm2/s at 15 Å |
Maximum field integral | 0.44 T·m |
Typical sample size |
3 cm × 3 cm |
Accessible scattering angle range |
from 2° to 105° |
Accessible Q range |
from 0.060 Å-1 to ≈1.80 Å-1 at λ = 5 Å |
| from 0.050 Å-1 to ≈1.50 Å-1 at λ = 6 Å |
| from 0.035 Å-1 to ≈1.10 Å-1 at λ = 8 Å |
| from 0.025 Å-1 to ≈0.80 Å-1 at λ = 11 Å |
| from 0.020 Å-1 to ≈0.60 Å-1 at λ = 15 Å |
Maximum field integral | 0.44 T·m |
Dynamic range |
from 3 ps to 10 ns at λ = 5 Å |
| from 5 ps to 15 ns at λ = 6 Å |
| from 10 ps to 40 ns at λ = 8 Å |
| from 50 ps to 100 ns at λ = 11 Å |
| from 150 ps to 200 ns at λ = 15 Å |
Detector | 32 cm × 32 cm PSD |