NG7 30-m Small Angle Neutron Scattering Instrument
Instrument Performance:
Flux at the Sample:
Flux (n/cm2-sec) at the sample versus wavelength (Å). The upper curve is for the maximum number of guides (Ng = 8), giving a source aperture size of 5.0 cm diameter; the lower curve is for no guides (Ng = 0) in the pre-sample collimation section and a source aperture of 3.8 cm diameter. The Δλ/λ was set at 11%.
'Typical' SANS Data:
![]()
![]()
SANS data for a 1 mass % solution of the protein Apoferritin (456 K daltons) in D2O. The data were collected at two sample-to-detector distances (1.6 m and 6.0 m) at a wavelength of 6 Å (11 % FWHM wavelength spread). Counting time was 15 min. at each distance. The data are fit (red line) to a spherical core/shell model smeared (convoluted) with the instrumental resolution function.
NG7 Home | NG7 Specifications | NG7 Performance | NG7 Publications
NG7 Schedule | NG7 Status | SANS Home | SANS Equipment
Last modified 29-July-2005