NG3 30-m Small Angle Neutron Scattering Instrument
Characteristics and Performance:
- Source: neutron guide (NG-3), 60 mm x 60 mm
- Monochromator: mechanical velocity selector with variable speed and pitch
- Wavelength Range: 0.5 nm-2.0 nm
- Wavelength Resolution: 10%-30% (FWHM)
- Source-to-Sample Dist.: 4 m to 16 m in steps via insertion of neutron guide sections
- Sample-to-Detector Dist.: 1.3 m to 13 m
- Collimation: circular pinhole collimation
- Sample Size: 0 to 25 mm diam
- Q-Range: 0.015 nm-1 to 6 nm-1
- Detector: 650 mm x 650 mm 3He position-sensitive proportional counter (5 mm resol.)
![]()
Neutron Flux on Sample as a Function of Qmin*:
![]()
Qmin (nm-1)
Is(n/sec) in units of 105
0.015
0.2
0.020
0.5
0.040
4.0
0.1
30
*Measurements taken on a 15 mm diameter sample, and Δλ/λ = 0.15 (FWHM).
NG3 Home | NG3 Characteristics & Performance | NG3 Description | NG3 Publications
NG3 Schedule | NG3 Status | SANS Home | SANS Equipment
Last modified 18-October-2005