Neutron spin filters can be applied to measurements requiring different field orientations. The schematic shown above illustrates the use of NSFs on MAGIK. In this example, the NSF is the analyzer for a polarized beam measurement with a horizontal field. For more details concerning the neutron spin filter, please select the provided links for general {Chen03} and specific {Chen04} applications. For more info concerning the MAGIK instrument, please contact Brian Maranville (contact info below).



Specifications for Neutron Spin Filter on MAGIK

NSF on ANDR
  • 3He cell analyzer only
  • 3He polarization flipping capability
  • Wavelengths: λ = 5 Å
  • Measurable Q range: covers most of 2D PSD
  • Sample field: H ≤ 0.7 Tesla
  • Four cross-section polarization correction
  • 3He transmission: ≤ 45% (for desired spin state)
  • Flipping ratio: ≤ 50