HFBS OC
OC-HFBS (24K) detailed information about orange cryostats

The HFBS Orange Cryostat was specifically designed for use on the HFBS instrument at the NCNR. It has a modified tail set that fits into the HFBS pit and a modified body that has a large flange to support its weight. There is only one HFBS cryostat available and it has a 50mm sample well diameter.

Useful Dimensions (approximate)
System DCS HFBS 100mm 70mm 50mm
A 1358 1304 1192.5 942 974
B 170 152 152 152 70
C 353 400 431 342 127
D 325 380 395 330 330
E 1010 910 800 565 565
F 263 310 355 306 230
G 1330 1265 1096 910 910
System Preparation
Pump out Vacuum Jacket > 2 hrs.
LHe Fill ~30 min. / ~30 L
Cooldown to Base Temp. ~2 hrs.
*Click Here for Sample Mounting Info*


Other Capabilities

The capabilities stated so far have been for operation in the standard setup. When using a special setup these capabilities may change, however, many special setups have been used with success. Some of these include: gas handling, transport measurements, high pressure, etc. Any special setups should be discussed with the sample environment staff before being scheduled.